Models 1600 & 1630 Oxygen Controllers
|
|
|
Models 1600 and 1630 provide economical atmospheric and dissolved oxygen control. Each has the latest in microprocessor-based technology with an unprecedented number of standard features found in a 1/4 DIN package. All control functions are conveniently programmable from the front panel.
Model 1600's galvanic electrode measures up to 100. parts per million (ppm) dissolved oxygen in aqueous solutions as well as up to 100% concentration in an atmospheric environment. Model 1630's galvanic electrode measures 0-100% atmospheric oxygen.
There are no filling solutions or membranes to change, making the electrodes of both controllers virtually maintenance-free. Models 1600 and 1630 come with a standard 5 ft. oxygen electrode. Custom lengths can be manufactured to 100 ft.
The electrodes are temperature compensated and are totally sealed from the environment. Model 1600's electrode features a low flow-rate sensitivity, needing only 0.5 ft/second flow to read dissolved oxygen accurately. No flow is needed when testing atmospheric oxygen
Standard features found in Models 1600 and 1630 include:
|
|
Specifications
| Model 1600 | Model 1630 | |
| Range: |
0-100 ppm Dissolved Oxygen 0-100% Atmospheric |
0-100% Atmospheric |
| Accuracy: | ± 0.25 ppm, ± 1% of full scale | ± 1% of full scale |
| Resolution/Precision: | 0.1/ (±) 0.1 | 0.1 / ± 0.1 |
| Response Time: | 90% in 12 seconds | 95% in 30 seconds |
| Electrode Warranty: | Six Months | One Year |
| Electrode Dimension: | 1.62"L x 1.25" Dia. | 2" L x 1" Dia. |
| Optional Outputs: | 4-20mA & RS232 | 4-20mA & RS232 |
Zero: True zero in the
absence of oxygen.
Temperature Compensation: Automatic from 0-40°C
Electrode: Standard 5 ft. length, with 0.25 inch phone plug
Set Point Range: Selectable
Displays: Two 4 digit, 7 segment 0.3" high LED's: PV red, SV green
Supply Voltage: 85 to 265 VDC or VAC, 50 through 400 Hz. This
applies to the instrument power only. Consult the factory for relay and
SSR output operation at DC and power line frequencies other than 50/60 Hz.
Power Consumption: 5 VA maximum
Operating Temperature Range: -10 to +55°C, controller; 0 to 40°C,
electrode
Storage Temperature Range: -40 to +80°C, controller; refrigerate
electrode
Humidity Conditions: 0 to 90% up to 40°C non-condensing; 10 to 50%
at 55°C non-condensing
Memory Backup: Non-volatile memory, no batteries required
Control Output Ratings: SSR: 3.5A @ 25°C typical; derates to 1.25A @ 55°C
Panel Cutput: 3.58" x 3.58"; 9.9 x 9.9 cm
Depth Behind Mounting Surface: 7 inches, 17cm
Weight: 2.8 lbs; 1.27 kg
Controller Warranty: 2 Years
NOTE: These controllers are not intended for use in life support
systems, nor are the controllers 100% fail-safe in some modes of operation.
Please contact the manufacturer to determine the suitability of the controllers
for your application.
| Item | Model/Description |
Price |
| 01-1600-Control | Model 1600 Oxygen Controller |
$962.00 |
| 01-1600-CONT420 | Model 1600 Oxygen Controller w/4-20mA Output |
1,023.00 |
| 01-1630-Control | Model 1630 Oxygen Controller |
1,300.00 |
| 01-1630-CONT420 | Model 1630 Oxygen Controller w/4-20mA Output |
1,362.00 |
| 01-OX630-Elec | Replacement Electrode |
521.00 |
Search | Contact | Another Link
![]()
All shipments made within the state of Texas are subject to Texas sales tax. If your firm is exempt from sales tax, please submit proper tax exemption certificate with your order.
Instructions for Companies outside of the United States: Terms are Wire Transfer of funds in US dollars to our bank to cover the cost of merchandise purchased plus shipping charges. Bank information provided upon receipt of your firm Purchase Order.